After additional information from the customer, the reported event was further clarified to reflect that the malfunction was not related to image correlation.The reported event was related to a loss of system function of the device due to the workflow not being followed.The interruption of the workflow of the device and software application being used for the procedure would not allow the initialization of correlation, therefore lack of accuracy due to correlation would not be possible.As this event does not relate to a correlation inaccuracy, it is not likely to contribute to or cause a serious injury and was not a reportable event per cfr 21, part 803.This supplemental is being filed to identify a mdr was not required for this event.
|