Catalog Number 10697306
Device Problem Use of Device Problem (1670)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/07/2018
Event Type  malfunction  
Manufacturer Narrative
Siemens customer support is requesting to find out if there is evidence of benzalkonium interference on this measurement cartridge to understand the discrepancy of the sodium results.From the data that the customer provided, there were several indicators of sample quality issues: d39 obstruction - clot detected, question results, out of range results and insufficient sample.After the error message: sodium sensor interferent detected, the sensor became very unstable.There were many benzal sample error messages when patient samples were run.As per the rp500 operators manual, benzalkonium is a known interfering substance to the na sensor and should be avoided.
 
Event Description
The customer reported discrepant sodium results on two samples run on the rp 500.There is no reported injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
steven andberg
2 edgewater drive
norwood, MA 02062
7812693655
MDR Report Key7883076
MDR Text Key120709933
Report Number3002637618-2018-00108
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K122539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10697306
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/24/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-