• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD ENDO CLIP II; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD ENDO CLIP II; CLIP, IMPLANTABLE Back to Search Results
Model Number 176657
Device Problems Mechanics Altered (2984); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/22/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during a prostatectomy, when the surgeon needed to clip a vessel to release the prostate, the surgeon used the first clip applier, there was a snap at first use then it did not clip.When he used the second clip applier to replace the first defected one, the clip closes badly / the yellow safety is not fixed on the device.So they used a third clip applier at the request of the surgeon.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.There were no visual abnormalities noted during the evaluation.The instrument was engaged in the end of firing safety interlock.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP II
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
MDR Report Key7887190
MDR Text Key120696492
Report Number9612501-2018-01667
Device Sequence Number1
Product Code FZP
UDI-Device Identifier20884521057866
UDI-Public20884521057866
Combination Product (y/n)N
PMA/PMN Number
K954435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 03/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/18/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2022
Device Model Number176657
Device Catalogue Number176657
Device Lot NumberJ7F2147CX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2019
Date Manufacturer Received02/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-