Patient age: 18 years or older.Device evaluated by manufacturer - visual inspection revealed that the distal end had a slight left curve after the steering knob was returned to neutral.Dried body fluid was found under the edges of ring 1 & 3 electrodes.Dried body fluid also found on the handle, main body, distal end and inside the luer fitting/tubing.Dried saline was found on the distal end and inside the irrigation ports.Electrical continuity checks revealed no electrical opens or shorts, as checked manually using a multi-meter and breakout box.All electrodes, sensor and thermocouple resistances measured in spec and were typical.The steering knob and the tension control knob functioned properly on both lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Connected device to a maestro 4000 to measure tip temperature.During 5 minutes at room temperature (22.0c), the maestro displayed 22c.After soaking 5 minutes in a tank of 37.0c saline, the maestro displayed 37c.Ablation was verified by using a maestro generator 4000, a metriq pump, and a tank of 37c saline solution.Occlusion errors occurred on the pump display, so ablation testing was performed on the bench without irrigation connected.The device was found within specifications.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|