• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS PRODUCT GMBH COAGULATION FACTOR IX DEFICIENT PLASMA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS PRODUCT GMBH COAGULATION FACTOR IX DEFICIENT PLASMA Back to Search Results
Model Number COAGULATION FACTOR IX DEFICIENT PLASMA
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/18/2017
Event Type  malfunction  
Manufacturer Narrative
The customer contacted a siemens customer care center (ccc) and reported that discordant, falsely elevated coagulation factor ix deficient plasma (fix) results were obtained on 4 samples from the same patient on the bcs xp system.There was no indication of a system malfunction and there were no reports of hardware errors when the discordant results were obtained on the patient samples.Quality controls were within expected ranges.Siemens is investigating the issue.Mdr 9610806-2018-00095 and mdr 9610806-2018-00096 were filed for the discordant results obtained on (b)(6) 2017 and (b)(6) 2017 for the same issue.Mdr 9610806-2018-00097 and mdr 9610806-2018-00098 were filed for discordant results obtained on samples from the same patient on (b)(6) 2018 and (b)(6) 2018, for the same issue.
 
Event Description
Discordant, falsely elevated coagulation factor ix (factor ix) results were obtained on a patient sample on a bcs xp instrument, using the coagulation factor ix deficient plasma reagent.The sample was repeated using a non-siemens methodology, resulting lower.On (b)(6) 2017, the sample was repeated using the initial reagent lot, resulting in discordant, falsely elevated factor ix results.The patient blood was redrawn and the new sample was run multiple times, using the same reagent lot on (b)(6) 2017 and (b)(6) 2017, resulting in other discordant, falsely elevated factor ix results.The redrawn sample was also repeated using a non-siemens methodology, resulting lower.The customer indicated that some of the results determined by the coagulation factor ix deficient plasma reagent were reported to the physician(s), who questioned the results.The customer also stated that correct results were reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated factor ix results.
 
Manufacturer Narrative
Siemens filed the initial mdr 9610806-2018-00094 on 19-september-2018.Additional information (20-september-2018): the customer clarified that the results obtained on (b)(6) 2017 were used for troubleshooting purposes.Siemens determined that the instruments and reagents used within the different tests and quality controls (qcs) were acceptable.Siemens determined that the genetic disorder of the patient potentially contributed to the tests to react differently.Corrected information (21-september-2018): the affected factor ix (fix) lot is 504122b.This information was corrected in the lot #.The year was revised for sample id (b)(6).On "(b)(6) 20117" was corrected to "(b)(6) 2017".The reagent and instrument are performing according to specifications.No further evaluation of this device is required.Mdrs 9610806-2018-00095_s1, 9610806-2018-00096_s1, 9610806-2018-00097_s1, and 9610806-2018-00098_s1 were filed for the same event.
 
Manufacturer Narrative
Additional information (24-october-2018): on 24-oct-2018, siemens was notified that customer reported the event to the fda.The customer also indicated that the patient has history of bleeding after hypospadias repair and a tonsillectomy at age (b)(6).The report filed by the customer indicates that the event date was for (b)(6) 2017, however the results correspond to the data dated (b)(6) 2018 in the initial mdr, in which was reported to siemens by the customer.Mdrs 9610806-2018-00095_s2, 9610806-2018-00096_s2, 9610806-2018-00097_s2, and 9610806-2018-00098_s2 were filed for the same event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGULATION FACTOR IX DEFICIENT PLASMA
Type of Device
COAGULATION FACTOR IX DEFICIENT PLASMA
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS PRODUCT GMBH
emil-von-behring-str. 76
marburg, D-350 41
GM  D-35041
MDR Report Key7890957
MDR Text Key120834034
Report Number9610806-2018-00094
Device Sequence Number1
Product Code GJT
UDI-Device Identifier00842768011023
UDI-Public00842768011023
Combination Product (y/n)N
PMA/PMN Number
K924396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup,Followup
Report Date 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/19/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/12/2018
Device Model NumberCOAGULATION FACTOR IX DEFICIENT PLASMA
Device Lot Number504122B
Was Device Available for Evaluation? No
Date Manufacturer Received10/24/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age16 YR
-
-