Model Number 302-20
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/28/2018
Event Type  malfunction  
Event Description
During replacement surgery, the new generator was connected to the old lead and high impedance greater than 10,000 ohms was seen.Generator diagnostics were performed with the new m105 and it was seen to be functioning properly.Pin re-insertion with the new generator and old lead was also attempted and did not resolve the issue.Lead replacement has not occurred to date.No additional or relevant information has been received to date.
 
Event Description
The patient underwent a lead replacement due to high impedance.The explanted lead has not been received for analysis to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7894108
MDR Text Key120973588
Report Number1644487-2018-01654
Device Sequence Number1
Product Code MUZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/06/2008
Device Model Number302-20
Device Lot Number1169
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received10/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/07/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age34 YR
-
-