• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS ANTI-TSHR IMMUNOASSAY; SYSTEM, TEST, THYROID AUTOANTIBODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS ANTI-TSHR IMMUNOASSAY; SYSTEM, TEST, THYROID AUTOANTIBODY Back to Search Results
Catalog Number ASKU
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/31/2018
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The customer questioned results for 1 patient tested for elecsys tsh assay (tsh), elecsys ft4 ii assay (ft4 ii) and anti-tshr on a cobas 8000 e 602 module.The customer thinks the tsh and anti-tshr results are too high and the ft4 ii results are too low.This medwatch will cover anti-tshr.Refer to medwatch with patient identifier (b)(6) for information on the tsh results and medwatch with patient identifier (b)(6) for information on the ft4 ii results.Twelve days after birth: the patient underwent neonatal mass screening using mass spectrometry and the tsh result was 79 uiu/ml.The patient was tested on the e602 module with a tsh result of 2467 uiu/ml, an ft4 ii result of 0.2 ng/dl and an anti-tshr result of 4.52 iu/l.The patient was administered levothyroxine (lt4) orally due to hypothyroidism.The patient's mother has a normally functioning thyroid and the mother's anti-tshr result was "negative." twenty two days after birth: a sample from the patient was sent to an external laboratory using an unknown method and the thyroid stimulating antibody result was 266% and the tsh-stimulation blocking antibody result was 25.8%.Forty four days after birth: the patient was tested on the e602 module and the tsh result was 86 uiu/ml and the ft4 ii result was 2.29 ng/dl.Oral lt4 was increased; the customer stated the tsh result was not normal for the high ft4 ii result.One hundred twenty seven days after birth: the patient was tested on the e602 module with a tsh result of 51.4 uiu/ml, an ft4 ii result of 1.85 ng/dl and an anti-tshr result of 3.34 iu/l.The sample was repeated by the architect method and the tsh result was 33.4 uiu/ml.The patient had an anti-tshr result by an unknown method of [-] 26.1%.The results from the patient's "latest" sample on the e602 module was a tsh result of 9.69 uiu/ml, an ft4 ii result of 1.95 ng/dl and an anti-tshr result of 3.16 iu/l.The results from the e602 module were reported outside of the laboratory.There was no allegation that an adverse event occurred.The e602 module serial number was (b)(4).
 
Manufacturer Narrative
The sample from (b)(6) 2018 was tested by the yamasa method and the anti-tshr result was < 1.0 iu/l.
 
Manufacturer Narrative
The "latest" sample mentioned in the initial report was from (b)(6) 2018.The sample from (b)(6) 2018 was submitted for investigation where the sample was run twice on an e602 module with anti-tshr results of less than 0.500 iu/l.From the information provided, a general reagent issue can be excluded.Values generated using different technologies cannot be directly compared to each other.Assays from different manufacturers, in this case abbott, can generate different results.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS ANTI-TSHR IMMUNOASSAY
Type of Device
SYSTEM, TEST, THYROID AUTOANTIBODY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7897358
MDR Text Key120983264
Report Number1823260-2018-03219
Device Sequence Number1
Product Code JZO
Combination Product (y/n)N
PMA/PMN Number
ASKU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup,Followup
Report Date 11/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/31/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age12 DA
-
-