BAXTER HEALTHCARE - MOUNTAIN HOME HOMECHOICE AUTOMATED PD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number R5C4479T
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/31/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the sterile packaging of eight (8) homechoice cassettes was damaged.The damage was described as the sterile packaging being torn open.This was noticed before use.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual devices were not available; however, photographs of the samples were provided for evaluation.Visual inspection of the photographs were performed which noted torn over pouches on five (5) samples and damaged master cartons.The reported issue was verified.The cause of the condition was undetermined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.A device analysis could not be completed on all samples as photographs were not received for every sample.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - MOUNTAIN HOME
mountain home AR
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7901733
MDR Text Key121587429
Report Number1416980-2018-06106
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue NumberR5C4479T
Device Lot NumberH18B05047
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received10/24/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-