Model Number 801188 |
Device Problems
Gas Output Problem (1266); No Flow (2991)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/30/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) was unable to duplicate the reported complaint.The epgs calibrated successfully and no errors were observed.The unit operated to the manufacturer's specifications.Per data log analysis, on (b)(6) 2018: 06:57:20 am the gas system is successfully calibrated.08:25:28 flow set to 1.35 liters per minute (l/min); 08:59:55 flow set to 2.99 l/min; 10:15:57 flow set to 1.99 l/min; 11:58:13 the perfusion screen is exited.There is no indication of a problem in the log.
|
|
Event Description
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) had no gas or oxygen (o2) flow to the system.The unit was restarted and the flow of gas and o2 returned.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The reported complaint could not be confirmed.Per the field service representative (fsr), there are no parts to return as the electronic patient gas system (epgs) was operating to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|