Model Number 801188
Device Problems Gas Output Problem (1266); No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/30/2018
Event Type  malfunction  
Manufacturer Narrative
The field service representative (fsr) was unable to duplicate the reported complaint.The epgs calibrated successfully and no errors were observed.The unit operated to the manufacturer's specifications.Per data log analysis, on (b)(6) 2018: 06:57:20 am the gas system is successfully calibrated.08:25:28 flow set to 1.35 liters per minute (l/min); 08:59:55 flow set to 2.99 l/min; 10:15:57 flow set to 1.99 l/min; 11:58:13 the perfusion screen is exited.There is no indication of a problem in the log.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the electronic patient gas system (epgs) had no gas or oxygen (o2) flow to the system.The unit was restarted and the flow of gas and o2 returned.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
The reported complaint could not be confirmed.Per the field service representative (fsr), there are no parts to return as the electronic patient gas system (epgs) was operating to the manufacturer's specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7903433
MDR Text Key121564498
Report Number1828100-2018-00493
Device Sequence Number1
Product Code DTQ
UDI-Device Identifier00886799000588
UDI-Public(01)00886799000588(11)060815
Combination Product (y/n)N
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801188
Device Catalogue Number801188
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-