MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMPREHENSIVE PRIMARY STEM 12MM MICRO; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMPREHENSIVE PRIMARY STEM 12MM MICRO; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Malposition of Device (2616)
Patient Problem Joint Dislocation (2374)
Event Date 08/29/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: catalog # lot # 113032 837990 versa-dial 42x18x46 hum head, cp0001156 076370 mini convertible e1 liner, 115395 413870 comp rvs cntrl 6.5x25mm st/rst, 118001 256620 versa-dial/comp ti std taper, 180552 651210 comp lk scr 3.5hex 4.75x25 st, 180553 571910 comp lk scr 3.5hex 4.75x30 st, 180556 723040 comp lk scr 3.5hex 4.75x45 st, pm0001410 676870 nerney lt pm mini conv glenoid, pm0001411 041070 nerney lt pm cnv accessory kit.Foreign: (b)(6).The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2018 -08923.Patient retained implants.
 
Event Description
It was reported that a patient underwent a total shoulder arthroplasty.Subsequently, the patient was revised less than three (3) months post-implantation due to dislocation.No further information is available.
 
Manufacturer Narrative
Patient medical records and investigative inputs confirmed the reported malposition of the humeral stem which caused the patient dislocation.The surgeon confirmed the initial version of the humeral stem was incorrect which led to the recurrent dislocations.Dhr was reviewed and no discrepancies were found.Root cause is attributed to unintentional user error in primary device positioning.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported the patient was revised to address chronic full and partial dislocation attributed to malposition of the humeral stem.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE PRIMARY STEM 12MM MICRO
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7906317
MDR Text Key121610782
Report Number0001825034-2018-08924
Device Sequence Number1
Product Code MBF
Combination Product (y/n)N
PMA/PMN Number
PK060692
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number113612
Device Lot Number948300
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/05/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-