• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM; FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM Back to Search Results
Model Number SDM-00001-2D
Device Problem Unintended System Motion (1430)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/11/2018
Event Type  malfunction  
Event Description
It was reported that during a biopsy exam the c-arm rotated uncommanded.No injury reported.The system logs were reviewed by technical support and a field engineer was dispatched to the site.It was determined that the c-arm switches needed to be replaced.Once this was completed the system was working as intended.It was also noted that compression force used was too low to disable motions and the customer was notified to use slightly more compression force to enable lock out.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELENIA DIMENSIONS MAMMOGRAPHY SYSTEM
Type of Device
FULL FIELD DIGITAL MAMMOGRAPHY SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury CT 06810
Manufacturer Contact
kristin fornieri
36 & 37 apple ridge road
danbury, CT 06810
2037318491
MDR Report Key7907737
MDR Text Key121713263
Report Number1220984-2018-00158
Device Sequence Number1
Product Code MUE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSDM-00001-2D
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-