Model Number 550
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/08/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A physician reported an error occurred on the laser resulting in the need to restart the machine.The gantry moved down during this process requiring the surgeon to change the level of the bed to protect the patients eye from the laser.After restarting the laser the machine was used as normal.Additional information requested.
 
Manufacturer Narrative
The root cause cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENSX LASER SYSTEM
Type of Device
OPHTHALMIC FEMTOSECOND LASER
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7915527
MDR Text Key122154857
Report Number2028159-2018-02049
Device Sequence Number1
Product Code OOE
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K163551
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number550
Device Catalogue Number8065998162
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/26/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age75 YR
-
-