Catalog Number 03L80-31
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Note: this information was previously submitted under manufacturer report 1628664-2018-01887.The likely cause was changed from the architect c8000 analyzer ln 01g06-11 sn (b)(4) to the carbon dioxide reagents ln 03l80-31 lot 52577uq02 on 14september2018.Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of field data, and a review of labeling.No adverse trend was identified for the customer's issue.No return patient sample was available.Historical performance of the reagent lot was evaluated using world wide data.The patient data was analyzed and within established control limits.Labeling was reviewed and found to be adequate.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Event Description
The customer observed a falsely elevated carbon dioxide result on the architect c8000 analyzer.The following data was provided: (b)(4) initial co2 32, repeat 29 mmol/l.There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINICAL CHEMISTRY CARBON DIOXIDE
Type of Device
CARBON DIOXIDE
Manufacturer (Section D)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer (Section G)
ABBOTT MANUFACTURING INC
1921 hurd drive
irving TX 75038
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key7919998
MDR Text Key122641664
Report Number1628664-2018-01923
Device Sequence Number1
Product Code KHS
UDI-Device Identifier00380740005962
UDI-Public00380740005962
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060295
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Catalogue Number03L80-31
Device Lot Number52577UQ02
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/28/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-