Model Number 8011-0501-01 |
Device Problem
Failure to Deliver Shock/Stimulation (1133)
|
Patient Problem
No Information (3190)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the associated defibrillator failed to discharge using these internal paddles.Complainant indicated that the clinician obtained another set of paddles to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
|
Manufacturer Narrative
|
The internal handles were returned to zoll medical corporation; the malfunction was duplicated and attributed to a defective discharge button on the paddle.The internal handles were deemed unrepairable.Therefore, replacement internal handles were sent to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|