MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION AUTOCLAVABLE INTERNAL HANDLES; INTERNAL PADDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION AUTOCLAVABLE INTERNAL HANDLES; INTERNAL PADDLE Back to Search Results
Model Number 8011-0501-01
Device Problem Failure to Deliver Shock/Stimulation (1133)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
 
Event Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the associated defibrillator failed to discharge using these internal paddles.Complainant indicated that the clinician obtained another set of paddles to continue treating the patient.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The internal handles were returned to zoll medical corporation; the malfunction was duplicated and attributed to a defective discharge button on the paddle.The internal handles were deemed unrepairable.Therefore, replacement internal handles were sent to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCLAVABLE INTERNAL HANDLES
Type of Device
INTERNAL PADDLE
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key7927199
MDR Text Key122527978
Report Number1220908-2018-02816
Device Sequence Number1
Product Code LDD
UDI-Device Identifier00847946022433
UDI-Public00847946022433
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K963781
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number8011-0501-01
Device Catalogue Number8011050101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/01/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-