MAUDE Adverse Event Report: TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH CORPORATION TOSOH HLC-723G8 ANALYZER G8 Back to Search Results
Model Number G8
Device Problem High Readings (2459)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/18/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Fse arrived onsite to address the reported event.Upon review of the customer's calibration and qc results, fse noted a slight shift.Fse confirmed the column count to be 2390 injections.Fse replaced the column, recalibrated and ran qc.No further issues were noted.No further action was required by field service.A 13 month complaint history review and service history review for similar complaints was performed for the serial number (b)(4) from 18aug2017 through aware date (b)(6) 2018.There were no similar complaints identified during the search period.The g8 operator's manual under chapter 3, assay operations, was reviewed: 3.5 daily check be sure to check the following items before starting an assay (start command).The hemoglobin a1c st aia-pack hba1c package insert states the following: quality control: in order to monitor and evaluate the precision of the analytical performance, it is recommended that aia-pack hba1c control set be assayed daily.The minimum recommendations for the frequency of running control material are: after calibration, controls are run in order to confirm the calibration curve.Controls should be run if certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, controls should be run in order to verify the overall performance of the tosoh aia system analyzer.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The most probable cause of the reported event was due to failure of the column.
 
Event Description
The customer reported out of range high level 2 quality control (qc) with lot number 7076 on their g8 analyzer.The customer recalibrated and reran qc, which repeated with high level 2 result.The customer was not able to get level 2 of qc within acceptable range even after recalibrating the g8 analyzer with a new lot of calibrator.The customer requested service in order to further investigate the out of range qc results.A field service engineer (fse) was dispatched to address the reported event, which resulted in delay in reporting of patient results for hemoglobin a1c (hba1c).There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOSOH HLC-723G8 ANALYZER G8
Type of Device
G8
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 105-8 623
JA  105-8623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586 23
JA   1058623
Manufacturer Contact
doria esquivel
6000 shoreline court
suite 101
south san francisco, 
MDR Report Key7932488
MDR Text Key123394593
Report Number8031673-2018-00913
Device Sequence Number1
Product Code LCP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131580
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/03/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG8
Device Catalogue Number021560
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA10/03/2018
Distributor Facility Aware Date09/18/2018
Device Age4 YR
Event Location Outpatient Diagnostic Facility
Date Report to Manufacturer10/03/2018
Date Manufacturer Received09/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-