• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOME AC POWER SUPPLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM HOME AC POWER SUPPLY Back to Search Results
Catalog Number 295600-001
Device Problem Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/20/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the reported issue did not prevent the freedom driver from performing its life-sustaining functions.In addition, the freedom driver has a redundant power source of multiple onboard batteries.The freedom home ac power supply will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce (b)(4) initial.
 
Event Description
The freedom home ac power supply is a component that enables the freedom driver to be plugged into an external power source.The customer, a syncardia certified hospital, reported that the patient turned in a freedom home ac power supply with a chip in the connector.The customer also reported that the cord works but the patient wanted it replaced.The customer also reported that there was no adverse patient impact.
 
Manufacturer Narrative
The visual inspection confirmed the customer-reported issue of a broken white connector, while also identifying a missing foot on the ac power supply brick.The functional inspection found the observed damage to be solely cosmetic, as the ac power supply passed all other sections of testing.The root cause of the broken white connector could not be conclusively determined but is likely the result of either rough handling or wear from usage.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM HOME AC POWER SUPPLY
Type of Device
AC POWER SUPPLY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7934894
MDR Text Key122811273
Report Number3003761017-2018-00476
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 11/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295600-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/28/2018
Date Manufacturer Received09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age21 YR
-
-