OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22306D
Device Problems Break (1069); Fracture (1260); Material Fragmentation (1261); Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The suspect medical devices have not yet been returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical devices are returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during an unspecified therapeutic myoma resection procedure, the loop wire at the distal end of two hf resection electrodes broke off and fell inside the patient.However, no fragments remained inside the patient since they were reportedly retrieved.No further information was provided, but there was no report about an adverse event or patient injury.
 
Manufacturer Narrative
The suspect medical devices were not returned to olympus for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.Furthermore, a dhr review could not be performed since basic data of article identification (lot number) are missing.Instead a manufacturing and quality control review was performed for the last 24 months of production.There were no non-conformities or deviations regarding the described issue.The case will be closed from olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
MDR Report Key7941963
MDR Text Key125136603
Report Number9610773-2018-00090
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051689
UDI-Public14042761051689
Combination Product (y/n)N
PMA/PMN Number
K100275
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 12/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22306D
Device Catalogue NumberWA22306D
Was the Report Sent to FDA? No
Date Manufacturer Received12/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-