WAVELIGHT GMBH ALLEGRETTO WAVE EYE-Q EXCIMER LASER; OPHTHALMIC EXCIMER LASER SYSTEM Back to Search Results
Catalog Number 8065990739
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem No Code Available (3191)
Event Date 09/10/2018
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.No udi required due to this device was out of production prior to the september 24, 2014 udi regulation date.(b)(4).
 
Event Description
A customer reported the patient's hyperopic treatment was over corrected.No additional information available.
 
Manufacturer Narrative
The device history records (dhr) for the device was reviewed.No abnormalities that could have contributed to this event were found.The associated device was released based on acceptance criteria.A review of the technical service onsite history showed no abnormalities that could have contributed to this event: laser was successfully verified prior to and after the day of the treatment.The root cause could not be identified conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
Manufacturer (Section G)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM   91058
Manufacturer Contact
bryan blake
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8176152230
MDR Report Key7943006
MDR Text Key123051509
Report Number3003288808-2018-01643
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number8065990739
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/23/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age20 YR
-
-