Patient procedure began and surgeon requested the storz bipolar coagulation/ dissection device.The surgeon nor the operating room tech noted anything wrong with the device.When the cautery used, the surgeon and tech noted a small amount of smoke and a few small black fragments.The surgeon removed the device, he and the operating room tech noted a small area on the instrument where the device had a break in the insulation approximately 1 cm from the tip.The surgeon then examined the patient laparoscopically again and noted a through and through burn to the patients left fallopian tube.The instrument had been cleaned and sterilized appropriately according to manufactures instructions for use.The storz instrument had been inspected prior to sterilization using the insuscan and no defects were found.
|