Catalog Number 8065990739 |
Device Problems
Improper or Incorrect Procedure or Method (2017); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Visual Impairment (2138)
|
Event Date 09/13/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.No udi required due to this device was out of production prior to the september 24, 2014 udi regulation date.The manufacturer internal reference number is: (b)(4).
|
|
Event Description
|
A customer reported an under correction of a monovision hyperopic treatment.The patient reported poor vision in both eyes.No additional information available.There are two related reports for this patient.This report addresses the patient's right eye, and another manufacturer report will be filed for the fellow eye.
|
|
Manufacturer Narrative
|
The device history records (dhr) for the device were reviewed.The associated device was released based on acceptance criteria.A review of the technical service onsite history showed no abnormalities that could have contributed to this event: laser was successfully verified prior to and after treatment date.The root cause could not be identified conclusively.The manufacturer internal reference number is: (b)(4).
|
|