Results code 2: 213: the engineering evaluation stated the following: the entire deice was returned.There was a kink in distal shaft, upon which the endoprosthesis is mounted, near the transition.There was approximately 1 cm of distal shaft exposed near tip.The knob doesn¿t appear to have been pulled from hub since the deployment line is taut within the hub.There were no separations in the delivery catheter.There was no deployment of the endoprosthesis noted.Engineering evaluation conclusion is inconclusive as it relates to the event description.Based on the device examination performed, no manufacturing anomalies were identified.
|