• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10603
Device Problems Device Damaged Prior to Use (2284); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 10/09/2018
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.
 
Event Description
It was reported that stent damage occurred.While opening the device during preparation, a 3.00 x 20 synergy drug-eluting stent was noted to be defective with a wire sticking up.The procedure was completed with another of the same device.No known complications were reported.Synergy ii us mr 3.00 x 20 mm stent delivery system was returned for analysis.A visual examination of the stent found proximal stent damage, stent struts were pulled and lifted.The undamaged distal section stent.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along several locations of the hypotube shaft.This type of damage is consistent with excessive force that could have been applied on the delivery system.A visual and tactile examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Type of Device
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key7949091
MDR Text Key123734398
Report Number2134265-2018-61726
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier08714729840176
UDI-Public08714729840176
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/22/2020
Device Model Number10603
Device Catalogue Number10603
Device Lot Number0022218079
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/14/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-