• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 40MM G; PROSTHESIS HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 NEUTRAL ARCOMXL LNR 40MM G; PROSTHESIS HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Information (3190)
Event Date 09/11/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Additional concomitant medical products: g7 pps ltd acet shell 58g, pn 010000666, ln 6124028; g7 neutral arcomxl lnr 36mm g, pn 010000742, ln 3822626.Multiple mdr reports were filed for this event.Please see reports: 0001825034-2018-09557, 0001825034-2018-09558.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that liner would not seat properly into a cup after several attempts.Due to removing original cup and re-implanting different cup, cleaning and exposing tissue, multiple attempts at seating, and liners being difficult to insert, there was a 45 minute delay.Attempts have been made and additional information on the reported event is unavailable at this time.No further information is available.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was not confirmed.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.The complaint was not confirmed.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL ARCOMXL LNR 40MM G
Type of Device
PROSTHESIS HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7951505
MDR Text Key123224718
Report Number0001825034-2018-09559
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/11/2019
Device Model NumberN/A
Device Catalogue Number010000748
Device Lot Number3387335
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/02/2018
Was the Report Sent to FDA? No
Date Manufacturer Received02/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age60 YR
-
-