• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: REDAX SPA SMART DRAIN CH 10 W/NEEDLE; ROUND FLUTED DRAIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

REDAX SPA SMART DRAIN CH 10 W/NEEDLE; ROUND FLUTED DRAIN Back to Search Results
Model Number 24702
Device Problem Complete Blockage (1094)
Patient Problem Hematoma (1884)
Event Date 08/13/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4) distributor, (b)(4), has visited and met the health professionals involved in this case and the following is what they have reported: the main complaint is that the drain is too soft; they say that due to this when the drain bends, it kinks and become occluded which has led to a patient having an haematoma and consequently further surgical treatment.Similar complaint is reported with incident report (b)(6) from the same hospital.The first analysis has been conducted on the production documents of the involved lot: no deviations have been observed on the production and on the labeling and qc records.The lot is made of (b)(4) units; no stock is available, all the units have been distributed on the market.Complaint data base has been reviewed, as well as trend analysis to evaluate similar cases in the last 2 years: trend on specific issue on the product code 24702.No complaints recorded.Trend on all issues on specific lot no.F1801100.No complaints recorded.Trend on product family with similar issue.Complaint from (b)(6).Since the sample was not available for analysis, we have taken from our stock some units coming from other lots in order to performs technical verification.Redax drains have been validated and are in compliance with european standard en 13868 "catheters - test methods for kinking of single lumen catheters and medical tubing" and with en 1618 appendix e relative to the drain flow.However we have repeated the tests in order to evaluate kink resistance and the length of the tubing loop in the kinking and the flow value of smart drain ch 10 and we have compared these data with the ones obtained from equivalent drains of 2 competitors of same size.The results of the tests indicate that is difficult to perform kinking test according to the standard: it is more difficult when the drain size is more little.However the kink length is < 2 cm for all the samples tested.The length of kinking doesn't change between redax and competitors' drains being equal size; while the flow is upper in redax' drains instead competitors ones.The comparison with competitors has been done only with drain ch 10.No actions will be performed.The tube is in compliance with the international standards and no further complaints have been recorded on this product family relative to this specific issue.We have the same issue only in (b)(6).The lot complained has been produced with a silicone material in 70 sha, while currently redax has substituted 70 sha with an 80 sha material.However we found that hardness of the raw material (70 or 80 sha) doesn't influence the kinking resistance and the length of the loop is equivalent.].
 
Event Description
Suggestion that smart drain has blocked post operative diep procedure resulting in haematoma and flap failure.Pt required readmission to theatre for additional surgical intervention.Local review of other potential cases has flagged one other with another 2 in addition now being considered as potentially related to this and datix incidents will be completed.Representative from (b)(4) coming into the trust to discuss further.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMART DRAIN CH 10 W/NEEDLE
Type of Device
ROUND FLUTED DRAIN
Manufacturer (Section D)
REDAX SPA
via galileo galilei 18
poggio rusco, mantova 46025
IT  46025
Manufacturer (Section G)
REDAX SPA
via galileo galilei 18
poggio rusco, mantova 46025
IT   46025
Manufacturer Contact
daniela malavasi
via galileo galilei 18
poggio rusco, mantova 46025
IT   46025
MDR Report Key7959352
MDR Text Key124332856
Report Number3013058659-2018-00007
Device Sequence Number1
Product Code GCY
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date 10/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/22/2023
Device Model Number24702
Device Catalogue Number24702
Device Lot NumberF1801100
Was Device Available for Evaluation? No
Date Manufacturer Received09/14/2018
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-