Catalog Number 8065990739
Device Problem Power Conditioning Problem (1474)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.No udi required due to this device being out of production prior to the september 24, 2014 udi regulation date.Manufacturing record reviewed.No abnormalities that could have contributed to this event were found.(b)(4).
 
Event Description
A technician reported the laser voltage went to one hundred percent and the procedure was interrupted at ninety-one percent.Additional information has been requested.
 
Manufacturer Narrative
During the onsite visit the tm (territory manager) passivated laser with six gas fills and running laser to heat it up.The tm looked in the logfiles and saw the customer had not been doing their weekly gas fills and energy checks.The tm performed system verification per service installation record.The root cause is that the user did not perform weekly gas fills and energy checks.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLEGRETTO WAVE EYE-Q EXCIMER LASER
Type of Device
OPHTHALMIC EXCIMER LASER SYSTEM
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key7959830
MDR Text Key123802166
Report Number3003288808-2018-01662
Device Sequence Number1
Product Code LZS
Combination Product (y/n)N
PMA/PMN Number
P020050
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990739
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-