• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. MTO,SERVICE REP,FOOTSWITCH,EP-1,; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. MTO,SERVICE REP,FOOTSWITCH,EP-1,; ARTHROSCOPE Back to Search Results
Catalog Number 7205397S
Device Problems Melted (1385); Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/21/2018
Event Type  malfunction  
Manufacturer Narrative
Foreign zip code: (b)(6).
 
Event Description
It was reported that when the black shaver pedal was plugged into the d2 it caused the mdu to overpower and caused issues with the shaver blase tip, the shaver pedal was oscillating without being activated and essentially melted the saver blade hub causing the shaver to malfunction.As soon as the pedal was unplugged and ceased to be used this problem did not occur.
 
Manufacturer Narrative
The reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.No containment or corrective actions are recommended at this time.A review of the device history record shows there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MTO,SERVICE REP,FOOTSWITCH,EP-1,
Type of Device
ARTHROSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
jim gonzales
7000 west william cannon drive
austin, TX 78735
5123585706
MDR Report Key7965655
MDR Text Key123868025
Report Number1643264-2018-00781
Device Sequence Number1
Product Code HRX
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K892062
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 11/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7205397S
Was Device Available for Evaluation? No
Date Manufacturer Received11/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-