On (b)(6) 2018, a 21mm masters valve was selected.When the valve was implanted and rotated with the recommended valve rotator, the leaflet fractured into two pieces during the rotation of the valve.The valve and pieces were removed and a second 21mm masters valve was used to complete the procedure.The procedure was extended by half an hour, but the patient did not experience any adverse consequences and is stable and discharged.
|
The reported event of a fractured leaflet was confirmed.One leaflet was fractured and dislodged from the orifice, but was not returned to abbott in its entirety.One interior butterfly apex of the dislodged leaflet was chipped.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the dislodged leaflet or orifice damage.However, the damage was consistent with some external force applied to the valve which overstressed the carbon material.The cause of the reported event remains unknown.Please note, per the instructions for use, (b)(4), "using the valve holder/rotator and an sjm¿ valve holder handle, rotate the valve in situ to the desired position.The valve should rotate freely.If resistance is noted, the valve holder/rotator may not be properly seated in the valve, the valve may not be in the fully closed position, or the valve may be oversized.If the valve does not rotate freely, do not force valve rotation.".
|