• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE REPLICATOR PLATE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE REPLICATOR PLATE Back to Search Results
Model Number EQUINOX SQUARE TORQUE DEFINE SCREW DRIVE KIT
Device Problem Unintended Movement (3026)
Patient Problems Failure of Implant (1924); Pain (1994)
Event Date 10/01/2018
Event Type  Injury  
Manufacturer Narrative
Pending evaluation.
 
Event Description
Revision due to pain.Patient presented with a painful anatomic shoulder and xrays revealed pegs of the caged glenoid below the glenoid in the axilla region.Upon revision the entire glenoid was retrieved from this region with all pegs intact, but the cage left behind in the glenoid.Glenoid was removed and a reverse shoulder was done on the intact humeral stem.Patient is expected to do well.
 
Manufacturer Narrative
The revision reported was likely the result of deformation of the central cage and incomplete seating of the glenoid component due to the central cage contacting the scapular wall during impaction at the time of the original surgery, which led the central cage disassociating from the central peg of the glenoid component.Having a disassociated glenoid component likely contributed to the patient¿s reported pain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE REPLICATOR PLATE
Type of Device
REPLICATOR PLATE
Manufacturer (Section D)
EXACTECH, INC.
MDR Report Key7974090
MDR Text Key123996518
Report Number1038671-2018-00805
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberEQUINOX SQUARE TORQUE DEFINE SCREW DRIVE KIT
Device Catalogue Number300-20-02
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/12/2018
Date Manufacturer Received10/11/2018
Patient Sequence Number1
Patient Age77 YR
-
-