Model Number 0874010
Device Problems Nonstandard Device (1420); Overheating of Device (1437); Loss of Power (1475); Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/06/2018
Event Type  malfunction  
Manufacturer Narrative
The device was requested for return.An evaluation of the device confirmed that this complaint is related to an issue in which otherwise fully functioning pumps display a battery missing error notification, which can be caused by repetitive unplugging and plugging when the battery is fully charged.This situation can potentially lead to a therapy interruption.The user is notified of such a malfunction via an audible and visual notification and is instructed via the instructions for use to contact their healthcare provider in order to receive a replacement device.
 
Event Description
On (b)(6) 2018, the customer alleged to medela (b)(4) that the motion negative pressure wound therapy device alarmed, then shut down and was hot to touch while on a patient; though, there was no serious injury, the patient did not experience wound deterioration and the patient did not require consultation and/or treatment by a medical professional.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVIA MOTION+ ENDURE SINGLE
Type of Device
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
MEDELA AG
lattichstrasse 4b
baar zug 6341,
SZ 
Manufacturer (Section G)
MEDELA AG
lattichstrasse 4b
baar zug 6341,
SZ  
Manufacturer Contact
jan kloiber
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key7974779
MDR Text Key125115587
Report Number1419937-2018-00235
Device Sequence Number1
Product Code OMP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161128
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number0874010
Device Catalogue Number0874010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2018
Date Manufacturer Received10/15/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-