(b)(4).Manufacturing site evaluation: the instrument arrived in a decontaminated condition.A microscopic investigation of the working tip, especially of the fracture surface.The fracture surface shows the typically hints for a forced brittle fracture under bending load, further we noticed a crack in the shaft, starting from the fracture site.A measuring of the tip rest shows, that about 8 mm of the tip are missing.The manufacturing documents have been checked and found to be according to specification valid during the time of production.There are no further complaints with this lot at hand.The root cause for the problem is most probably usage related.A capa is not necessary.
|
It was reported by the healthcare professional to the company sales representative "during insertion of a cannulated pedicle screw, the tip of the guide wire broke off and remained in the bone." customer clarification update: the s4 system instruments were used for this surgery: leading component: fw258r s4 aiming device f/kirschner boring wire during insertion the tip of the sleeve of the aiming device broke off.Involved component: fw271r s4 trocar f/canulated screws.Additional information submitted: it was reported that no fragments were retrieved, they remained in situ.The patient was restored without any consequences with no additional medical intervention provided.
|