Catalog Number 363095
Device Problems Short Fill (1575); Volume Accuracy Problem (1675)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/27/2018
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported with the use of the paxgene® blood rna tube there was an issue with underfill.There was no report of injury or medical intervention.
 
Manufacturer Narrative
Investigation summary: bd received samples from the customer facility for investigation.The samples were evaluated and the customer's indicated failure mode for underfill with the incident lot was not observed.Additionally, retention samples were selected from bd inventory for evaluation/testing and upon completion, the issue relating to underfill was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for underill with the incident lot was not observed.Evaluation/testing of the retain samples was also conducted and underfill was observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.Root cause description: a capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.
 
Event Description
It was reported with the use of the paxgene® blood rna tube there was an issue with underfill.There was no report of injury or medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAXGENE® BLOOD RNA TUBE
Type of Device
RNA PREANALYTICAL SYSTEMS
Manufacturer (Section D)
BECTON, DICKINSON AND COMPANY (BD)
belliver way
belliver industrial estate
plymouth
MDR Report Key7981798
MDR Text Key124501689
Report Number9617032-2018-02912
Device Sequence Number1
Product Code NTW
Combination Product (y/n)N
PMA/PMN Number
DEN050003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 10/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2019
Device Catalogue Number363095
Device Lot Number8144657
Date Manufacturer Received09/27/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-