A mechanical valve was implanted by dr.And the atrium was closed.Once the patient's heart was restarted, it was discovered that the mechanical leaflets on the valve were faulty and would not pivot to allow blood flow.The patient's heart was re-arrested, the faulty valve was explanted and a new sjm 23mm regent mechanical valve was successfully implanted.The faulty valve was saved and returned to the manufacturer's representative.Manufacturer response for heart-valve, mechanical, sjm regent¿ (per site reporter).The representative was given the defective mechanical valve 2 weeks after explant.The response is unavailable at this time.
|