MAUDE Adverse Event Report: PENUMBRA, INC. INDIGO SYSTEM CAT5 ASPIRATION CATHETER; DXE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. INDIGO SYSTEM CAT5 ASPIRATION CATHETER; DXE Back to Search Results
Device Problems Device Damaged Prior to Use (2284); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 09/21/2018
Event Type  malfunction  
Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2018-02110.
 
Event Description
During preparation for a medical procedure, a hospital staff noticed that the distal ends of two indigo system aspiration catheter 6¿s (cat6's) were kinked upon removal from the packaging.It was also noted that the cat6¿s would not flush properly.The damage to the cat6's were found prior to use, therefore, the cat6¿s were not used in the procedure.The procedure was completed using a new cat6.
 
Manufacturer Narrative
Results: the returned cat5 had ovalizations approximately 126.0 cm and 128.0 cm from the hub.Conclusions: evaluation of the returned cat5 revealed a device with distal ovalizations, which confirmed the customer reported kink.If the device is forcefully gripped or pinched during preparation, damage such as an ovalization may occur.During functional testing, the cat5 was able to be flushed with solution; however, the ovalizations limited flow.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The product lot number was not provided; therefore, the manufacturing records could not be reviewed.This report is associated with mfr report number: 3005168196-2018-02110.
 
Event Description
During preparation for a medical procedure, a hospital staff member noticed that the distal ends of an indigo system cat5 aspiration catheter (cat5) and an indigo system aspiration catheter 6 (cat6) were kinked upon removal from the packaging.It was also noted that the cat5 and cat6 would not flush properly.The damage to the devices were found prior to use, therefore, the devices were not used in the procedure.The procedure was completed using a new cat6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INDIGO SYSTEM CAT5 ASPIRATION CATHETER
Type of Device
DXE
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7997259
MDR Text Key124976990
Report Number3005168196-2018-02109
Device Sequence Number1
Product Code DXE
Combination Product (y/n)Y
PMA/PMN Number
K160533
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 01/01/2005,09/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2018
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received02/01/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-