LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL OT ULTRA TEST STRIPS; GLUCOSE MONITORING SYS/KIT Back to Search Results
Model Number N/A
Device Problem Peeled/Delaminated (1454)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
These complaints were initially ruled out at the time of complaint receipt as during customer services troubleshooting, there was no indication to reasonably suggest that the product malfunctioned and there was also no allegation of any adverse events as a result of the reported issues.Complaints relating to the reported product(s) were evaluated.It was concluded that the number of complaints for the product(s) did not breach thresholds indicative of a systemic quality or risk issue.Lifescan (lfs) received and further evaluated product returns for theses complaints.The test strip samples received failed testing, the reported issue was confirmed confirmed.The investigation revealed that the test strips were damaged during the slit and vial process.This report is processed under exemption number e2005018.
 
Event Description
This report summarizes 2 malfunction events.A review of the events indicated that the ot ultra test strips experienced a peeling test strip issue.These reports were received from various sources.It is not known if the patients associated with this allegation were male or female and there is no age or weight data available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT ULTRA TEST STRIPS
Type of Device
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug, 6300
SZ  6300
Manufacturer (Section G)
LIFESCAN SCOTLAND
beechwood business park north
inverness, scotland IV2 3 ED
UK   IV2 3ED
Manufacturer Contact
goncalo sousa
beechwood business park north
inverness, scotland IV2 3-ED
UK   IV2 3ED
MDR Report Key7999662
MDR Text Key124982020
Report Number3008382007-2018-02835
Device Sequence Number1
Product Code NBW
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K062195
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 10/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/24/2018
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberN/A
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date10/01/2018
Date Manufacturer Received10/01/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-