• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 HI-WALL ARCOMXL LNR 36MM F M F; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 HI-WALL ARCOMXL LNR 36MM F M F; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Failure to Align (2522)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that g7 liner would not sit upon impaction into the matching sized g7 acetabular shell.Surgery was completed with a new implant.Attempts have been made and additional information on the reported event is unavailable.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
Udi- (b)(4).Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the liner identified poly string hanging from the locking feature.There were also other visible damages on the locking features as well as scallops.These damages were likely caused during an attempt to impact the liner with the shell and during removal.Dhr was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 HI-WALL ARCOMXL LNR 36MM F M F
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key8004475
MDR Text Key125280243
Report Number0001825034-2018-09409
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
PMA/PMN Number
K121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/13/2023
Device Model NumberN/A
Device Catalogue Number010000819
Device Lot Number6255321
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Was the Report Sent to FDA? No
Date Manufacturer Received02/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-