The log file of the affected device was provided for the investigation.It could be confirmed that the device performed multiple warmstarts on the reported date of event.Based on the error code df40.2000 the warmstarts occurred due to a missing power source.However, the reported device failure 83.0008 was not stored on the log file.Based on the log entries the batteries were checked onsite without finding any deviation and the power supply was replaced and sent to the manufacturer for a detailed analysis.No failure of the power supply could be detected during a 24 hour test in a laboratory savina 300 device.Therefore, the root cause for the missing power source could not be determined.The device reacted as specified to a deviation within the power source by initiating a software-controlled reboot (warmstart) in an attempt to solve the issue.A warmstart is accompanied by an audible and visible alarm of high priority and the pneumatic system opens which reduces airway pressure to ambient pressure allowing for spontaneous breathing.In case the deviation cannot be remedied by multiple warmstarts, the ifu advises the user to continue ventilation without delay using another independent ventilator.The user further reported occurrences of the device failure 43.0120 which could also be found in the log file in 2017 and has not occurred since.However, this device failure is not related to the reported event.
|