Model Number 500101-001
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/27/2018
Event Type  malfunction  
Manufacturer Narrative
Review of the customer provided photograph confirmed the spring was missing from the cpc connector.Clinicians and patients are trained to thread a wire tie through the thumb release tab of the cpc connectors to prevent inadvertent disconnection of the cannulae from the drivelines.It is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch.Although a definitive root cause for the displaced spring could not be determined, it is possible that a cpc connector spring can be displaced when the wire tie is threaded through, or removed from, the connector during or after a driver switch or during driver training.Syncardia has an open corrective and preventive action (capa) to address this issue with cpc connectors.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The customer, a syncardia certified hospital, reported that the spring in the cpc connector of the cannula was displaced.The customer also reported that staff manipulated the cpc connector and the spring was subsequently dislodged and lost.The customer also reported that the cpc connector was replaced at the hospital and there was no clinical impact on the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA 70CC TAH-T
Type of Device
BIVENTRICULAR REPLACEMENT DEVICE
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key8009794
MDR Text Key126083399
Report Number3003761017-2018-00485
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003008
UDI-Public(01)00858000003008
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model Number500101-001
Device Catalogue Number500101
Device Lot Number114196
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/01/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Age53 YR
-
-