ALCON LENSX, INC. LENSX LASER SYSTEM; OPHTHALMIC FEMTOSECOND LASER Back to Search Results
Model Number 550
Device Problem Material Fragmentation (1261)
Patient Problems Injury (2348); Device Embedded In Tissue or Plaque (3165)
Event Date 10/09/2018
Event Type  Injury  
Manufacturer Narrative
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.Based on assessment, the product met specifications at the time of release.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported a patient with an anterior capsule tear during hydrodissection during laser assisted cataract surgery.The patient had an anterior vitrectomy performed.A piece of nucleus segment was lost to the posterior segment.The patient will have a retinal referral for the removal of the nucleus segment and is scheduled to have a vitrectomy.
 
Manufacturer Narrative
The root cause of the reported event cannot be determined conclusively.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LENSX LASER SYSTEM
Type of Device
OPHTHALMIC FEMTOSECOND LASER
Manufacturer (Section D)
ALCON LENSX, INC.
33 journey
suite #175
aliso viejo CA 92658
MDR Report Key8018900
MDR Text Key125460334
Report Number2028159-2018-02269
Device Sequence Number1
Product Code OOE
Combination Product (y/n)N
PMA/PMN Number
K163551
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/30/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number550
Device Catalogue Number8065998162
Was Device Available for Evaluation? No
Date Manufacturer Received12/03/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age67 YR
-
-