Model Number LC358M |
Device Problem
Fire (1245)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The "date of event" was not provided.The device has not been made available for evaluation at this time.Should the device or further information become available, a follow-up report will then be issued.
|
|
Event Description
|
Alleges when pressing the hand control, the lights went out, and when pressing the button again, it started smoking and then caught on fire.
|
|
Manufacturer Narrative
|
The device was returned and evaluated.The evaluator saw no evidence of fire.The chair, including the hand control (exposed wires), appeared to have been subjected to abuse.
|
|
Event Description
|
Alleges when pressing the hand control, the lights went out, and when pressing the button again, it started smoking and then caught on fire.
|
|