OTTO BOCK HEALTHCARE PRODUCTS GMBH C-LEG KNEE JOINT; EXTERNAL ABOVE KNEE PROSTHESIS Back to Search Results
Model Number 3C98-2
Device Problem Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
During investigation, a soldering error inside the tube adapter was determined which led to a change of the sensor signal.This causes inappropriate stance phase control which caused the occurred event (the joint is not holding in standing mode, it is flexing).
 
Event Description
The joint is not holding in standing mode, it is flexing.And the joint is not responding to commands from remote control.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C-LEG KNEE JOINT
Type of Device
EXTERNAL ABOVE KNEE PROSTHESIS
Manufacturer (Section D)
OTTO BOCK HEALTHCARE PRODUCTS GMBH
brehmstrasse 16
vienna, 1110
AU  1110
Manufacturer Contact
reinhard wolkerstorfer
brehmstrasse 16
vienna, 1110
AU   1110
MDR Report Key8022838
MDR Text Key125646749
Report Number9615892-2018-00026
Device Sequence Number1
Product Code ISY
Combination Product (y/n)N
Reporter Country CodeRS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number3C98-2
Device Catalogue Number3C98-2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/15/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age34 YR
Patient Weight70
-
-