• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAROTID WALLSTENT; STENT, CAROTID Back to Search Results
Model Number 26605
Device Problems Activation, Positioning or Separation Problem (2906); Material Integrity Problem (2978)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/25/2018
Event Type  malfunction  
Event Description
It was reported that stent partial deployment occurred.The target lesion was located in the carotid vessel.A 10.0-37 carotid wall stent was advanced to treat the lesion.However, at the time of implantation, the proximal part of the release system was damaged, leaving only 40% of the stent's surface open.A diagnostic catheter was advanced to make an exchange of guide and the stent was released.The procedure was completed with this device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8024995
MDR Text Key125681149
Report Number2134265-2018-62462
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
P050019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 10/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/17/2020
Device Model Number26605
Device Catalogue Number26605
Device Lot Number0020781504
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/18/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-