MEDTRONIC NEUROMODULATION ENTERRA; INTESTINAL STIMULATOR Back to Search Results
Model Number 37800
Device Problem High impedance (1291)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/23/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
"nformation" was received from a healthcare professional (hcp).It was reported that the hcp was in the operating room and were seeing >800ohms for impedance; the electrode impedances were: c2 20000ohms, c3 20ohms, and 23, 746ohms.No patient symptoms were reported.No further complications were reported/anticipated.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a healthcare professional (hcp).It was reported that the hcp was in the operating room and were seeing >800ohms for impedance; the electrode impedances were: c2 20000ohms, c3 20000ohms, and 23, 746ohms.No patient symptoms were reported.No further complications were reported/anticipated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERRA
Type of Device
INTESTINAL STIMULATOR
Manufacturer (Section D)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer (Section G)
MEDTRONIC NEUROMODULATION
800 53rd ave ne
minneapolis MN 55421 1200
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8026996
MDR Text Key125771006
Report Number3007566237-2018-03188
Device Sequence Number1
Product Code LNQ
Combination Product (y/n)N
PMA/PMN Number
H990014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number37800
Device Catalogue Number37800
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-