Model Number MMT-723NAH
Device Problem Erratic or Intermittent Display (1182)
Patient Problem Hyperglycemia (1905)
Event Date 10/20/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump had flashing display.The customer¿s blood glucose level was 255 mg/dl at the time of the incident.The customer stated that troubleshooting was performed for the partial display.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Device display properly and no missing segment anomaly noted.No unexpected no delivery alarm noted during testing.Device passed self test, rewind test, basic occlusion test, occlusion test, prime test, excessive no delivery test and displacement test.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAH
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key8032080
MDR Text Key126140329
Report Number2032227-2018-54807
Device Sequence Number0
Product Code OYC
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2018
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAH
Device Catalogue NumberMMT-723NAH
Device Lot NumberB5723NAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/24/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received01/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age30 YR
Patient Weight70
-
-