AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-024
Device Problem Retraction Problem (1536)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/16/2018
Event Type  Injury  
Event Description
On (b)(6) 2018, a 24mm amplatzer septal occluder (lot number: 6184089) was deployed using a 10f amplatzer torqvue delivery system (lot number: 5790870).While repositioning the device, the la disc failed to be re-captured.The physician removed the device through the femoral vein and the case was aborted.The patient was taken to surgery to close the asd.The patient is reported to be stable.
 
Manufacturer Narrative
The reported event of failure to recapture the left atrial disc could not be confirmed.The results of this investigation confirmed the device all functional and dimensional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8032663
MDR Text Key125932796
Report Number2135147-2018-00198
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010199
UDI-Public00811806010199
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/02/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model Number9-ASD-024
Device Catalogue Number9-ASD-024
Device Lot Number6184089
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age38 YR
Patient Weight60
-
-