SMITH & NEPHEW, INC. ARTICULAR SURFACE PROVISIONAL SZ 5 8MM; INSTRUMENT, MANUAL, SURGICAL, GENERAL USE
|
Back to Search Results |
|
Catalog Number 00584102508 |
Device Problem
Material Deformation (2976)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/16/2018 |
Event Type
malfunction
|
Event Description
|
It was reported that total knee replacement was performed and during trailing stage, femoral trial gouged into trial insert.Debris from trial insert was removed from wound.No delay was reported.
|
|
Manufacturer Narrative
|
The associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device had numerous scratches, nicks and a large gouge on the face.The device was manufactured in 2012 and exhibits signs of significant wear/usage.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
|