CONVATEC DOMINICAN REPUBLIC INC NATURA; NOT APPLICABLE Back to Search Results
Model Number 402204
Device Problem Malposition of Device (2616)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Based on the available information, this event is deemed to be a reportable malfunction.An investigation was performed to determine the root cause associated with the investigation wafer disc decentralization, complaint malfunction (skin barrier starter hole is defective, e.G.Misalignment or off center, leakage may occur), for lots manufactured in (b)(4).The cause and effect investigation was performed.  it was determined that the following probable root causes and opportunities were identified: machinery is the root cause of this incident.  it is concluded that all the machinery/ tooling items complied when compared against drawing specifications, however due to the demands of the process, these tooling require a dimension modification to reduce the variability of the process.Observations include:  misalignment of the wafer loading pins, misalignment of the upper wafer loading plate and fixation of the upper wafer loading plate and lower wafer loading plate.Actual testing method of (random hourly sampling) may not effectively capture the defects prior to packaging, it is suggested the implementation of a continuous testing method to anticipate to all of the peaks.A certification of the operators who have direct influence due to the ¿flange/wafer loading¿ process in operations, to reduce the learning curve effect in process and guarantee the training effectiveness.This should be performed in conjunct with the quality inspector, process engineer of such line and the supervisor.Actions will be taken for each factor and will be summarized on a corrective action / preventive actions (capa) plan.The investigation associated with this event is approved and complete.  this issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number: (b)(4).
 
Event Description
It was reported hole is not centered correctly so the edges overlap with the sticky surface, which makes it very difficult to attach the wafer to the skin.No photo was provided.
 
Manufacturer Narrative
Common device name: protector, ostomy.Exe; exempt.A batch record review was carried out on 10/30/2018 for the affected lot number 8g01612.The affected lot number was manufactured following the applicable procedures, all the testing scheme comply with the applicable documentation, the line clearance was executed per standard operating procedures and the results were documented.The material complies with specifications and other procedures were revised.The results are satisfactory.No photo/ no sample is available for evaluation related to the reported problem.The purpose of this investigations was to determine the root cause associated investigation wafer disc decentralization, complaint malfunction code ost-pmc1.8 (skin barrier starter hole is defective, e.G.Misalignment or off center, leakage may occur), for lots manufactured in convex 2 pc building 8, haina, d.R.It was determined the following root causes and opportunities: due to the fact that the occurrence of this failure mode is not constant, it has peaks of occurrence during the process, it is observed that the actual testing method (random hourly sampling) may not captured effectively the defects prior to packaging, so it is suggested the implementation of a continuous testing method to anticipate to all of the peaks.A certification of the operators who have direct influence due to the ¿flange/wafer loading¿ process in the operation they are executing should be considered in order to reduce the learning curve effect in process and guarantee the training effectiveness.This should be performed in conjunct with the quality inspector, process engineer of such line and the supervisor.It is considered that due to the observations registered, machinery is the root cause of this incident.It is concluded that all the machinery/ tooling items complied when compared against drawing and procedural instruction specifications, however due to the demands of the process, these tooling require a dimension modification to reduce the variability of the process.The following observations are: misalignment of the wafer loading pins.Misalignment of the upper wafer loading plate.Fixation of the upper wafer loading plate and lower wafer loading plate.Actions will be taken for each factor and are going to be summarized on corrective action / preventive actions (capa) plan.Actions covered in this investigation will be implemented in convex 2 pc building 8 inside convatec d.R., except for automatic convex 2 pc, because the design and functioning of this machine is different.This issue will be monitored through the post market product monitoring review process.No additional information has been provided to date.Should additional information become available a follow-up report will be submitted.Reporting site: (b)(4).Manufacturing site: (b)(4).
 
Event Description
To date no additional patient or event details has been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NATURA
Type of Device
NOT APPLICABLE
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
MDR Report Key8037733
MDR Text Key126381004
Report Number9618003-2018-02580
Device Sequence Number1
Product Code EXD
Combination Product (y/n)N
PMA/PMN Number
UNK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date07/27/2023
Device Model Number402204
Device Lot Number8G01612
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received10/31/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-