• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GIVEN IMAGING LTD., YOQNEAM BRAVO; ELECTRODE, PH, STOMACH Back to Search Results
Model Number FGS-0450
Device Problems Signal Artifact/Noise (1036); No Display/Image (1183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/27/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, there were lots of gaps in the study wherein the signal was n/a which meant that the recorder did not receive data from the capsule for any reason.The problem was the signal interruptions which could be caused by malfunctioning of the capsule or interference, rather than high values of ph.There were around 45 hours of missing data (almost two days) on a total of four days.For this reason, the software only detected six acid refluxes, but there could be many more hiding in those 45 hours of missing data.The recorder worked correctly during the previous procedure.There was no patient and user harm and a repeat procedure was necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BRAVO
Type of Device
ELECTRODE, PH, STOMACH
Manufacturer (Section D)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer (Section G)
GIVEN IMAGING LTD., YOQNEAM
yetsira 13 street
yoqneam 20692
Manufacturer Contact
amy beeman
161 cheshire lane, suite 100
plymouth, MN 55441
7632104064
MDR Report Key8043080
MDR Text Key126359975
Report Number9710107-2018-01036
Device Sequence Number1
Product Code FFT
UDI-Device Identifier07290101363064
UDI-Public07290101363064
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K102543
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-0450
Device Catalogue NumberFGS-0450
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/11/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-