• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977A260
Device Problem Positioning Problem (3009)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2018
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The healthcare provider reported via the manufacturer representative that they could not steer the lead when attempting to place it for implant.The surgeon was unable to steer the first lead even after switching out the stylets.A second lead was opened and there was no trouble steering the second lead.The surgeon then tried to use the stylets from the second lead in the first lead but the steering was still very difficult.A third lead was opened and the surgeon had no difficulty steering the third lead.The issue was resolved at the time of the report.No patient symptoms reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Analysis of the lead (b)(4) found no anomaly.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VECTRIS SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8043426
MDR Text Key126514907
Report Number2649622-2018-18133
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00763000006501
UDI-Public00763000006501
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 12/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/04/2022
Device Model Number977A260
Device Catalogue Number977A260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
-
-