Model Number CD2357-40C
Device Problems Premature End-of-Life Indicator (1480); Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/22/2018
Event Type  Injury  
Manufacturer Narrative
The device is included in the battery performance alert advisory issued by abbott on 28 august 2017.
 
Event Description
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician.It was noted that the elective replacement indicator and end of service were triggered on (b)(6) 2018.The device was explanted and replaced on (b)(6) 2018.
 
Event Description
New information noted that the patient was fine before, during, and after the procedure.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8044601
MDR Text Key126345429
Report Number2017865-2018-16625
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734508063
UDI-Public05414734508063
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P910023
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 12/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2016
Device Model NumberCD2357-40C
Device Lot Number4602851
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/29/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received12/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0115-2017
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age89 YR
Patient Weight84
-
-