Results: the cat6 product box was creased in two locations.The returned cat6 had a kink approximately 123.0 cm from the hub.The device also had ovalizations approximately 124.0 cm, 131.0 cm and 134.0 cm from the hub.Conclusions: evaluation of the returned cat6 revealed the distal tip was damaged.If the product was mishandled during preparation, damage such as this may occur.Further evaluation revealed the product box was also damaged.If the device was stored or handled poorly, this device may also have become damaged within the packaging.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a medical procedure, the hospital staff noticed that the tip of an indigo system aspiration catheter 6 (cat6) was crushed upon opening the package.The damage to the cat6 was found prior to use and, therefore, it was not used in the procedure.The procedure was completed using a new cat6.
|